Panorama™ Operating Instructions 0070-00-0575-50 xv
WARNING: Do not put MPSO (Multiple Portable Socket Outlets, i.e.,
multiple outlet extension cords) used with the Panorama
Central Station System on the floor. Connect only Panorama
Central Station accessories and components to the same
MPSO as the Panorama Central Station. Do not overload
MPSOs. Use only MPSOs that comply with the requirements
of IEC 60601-1-1.
WARNING: The Panorama Central Station will not monitor any non-
lethal arrhythmia, numeric or technical alarms for the
selected patient tile when the Suspend All Non-lethal Alarms
button is enabled.
WARNING: Use of accessories and cables other than those approved by
Datascope may result in increased Emissions or decreased
Immunity of the Telepack.
WARNING: The Telepack should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the
Telepack should be observed to verify normal operation in
the configuration in which it will be used.
WARNING: Ensure that the ECG lead wires are neatly secured in such a
way to prevent the lead wires from encircling the patient’s
neck and causing possible strangulation.
WARNING: PACEMAKER PATIENTS - Indication of the heart-rate may be
adversely affected by cardiac pacemaker pulses or by
cardiac arrhythmias. Keep pacemaker patients under close
surveillance. See ‘‘ECG Performance Requirements’’ on page
15-2 for disclosure of the pacer rejection capability.