Panorama™ Operating Instructions 0070-00-0575-50 7 - 7
Views Bedside Tab
Horizontal Calipers (Amplitude)
1. Select the Horiz button. A dashed horizontal reference line is placed at the bottom of
the waveform. The dashed reference line represents the 0 value starting point.
2. Adjust the location of the dashed horizontal reference line as follows:
• For coarse adjustments, select a point in the general area of the desired location. The
reference line will move to that point.
• For fine adjustments, use the up/down arrow keys to position the reference line at the
desired point.
3. Once the dashed horizontal reference line has been adjusted to the desired location,
select the Horiz button again. The dashed horizontal reference line is locked at its
current location. A solid horizontal measurement line is placed on the waveform above
the dashed horizontal reference line.
4. Adjust the location of the solid horizontal measurement line, as necessary, in the same
manner described in step 2.
5. Once the solid horizontal measurement line has been adjusted to the desired location,
select the SET button. The solid horizontal measurement line is locked at its current
location. The value for amplitude is calculated and displayed above the Horiz button.
• If desired, a vertical measurement can be performed while maintaining the horizontal
measurement information.
• If another horizontal measurement is desired, select the Clear button and repeat this
procedure.
7.1.3.6 Print Sidebar Button
Select the Print sidebar button to provides an All Strips Report for the selected patient tile.
7.1.3.7 Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, close
all tabs, message boxes and dialog boxes.
7.1.4 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Bedside tab.
MESSAGE/ISSUE * REASON SOLUTION
A waveform parameter is
not displayed.
If the waveform parameter
exists, it has not been scrolled
into view on the display.
Use the View More sidebar button
to scroll to the desired parameter
waveform.
The waveform parameter is
not being monitored at the
bedside.
Connect the sensor(s) for the
waveform parameter to the patient.
A digital parameter is not
displayed.
If the digital parameter exists,
it has not been scrolled into
view on the display.
Use the View More sidebar button
to scroll to the desired digital
parameter.
The digital parameter is not
being monitored at the
bedside.
Connect the sensor(s) for the digital
parameter to the patient.
* Messages are shown in all bold text.