User's Manual

Bedside Tab Views
7 - 4 0070-00-0575-50 Panorama™ Operating Instructions
One lead of ECG will always occupy the top waveform position, and will not scroll. This
ECG waveform is the same waveform shown on the main screen. If a view of a different ECG
lead is desired, it must be selected from the Display tab. See ‘‘Display Tab’’ on page 5-29
for additional information. All other waveforms can be scrolled into viewable position. The
scrolling feature provides access to a wrapping list that returns to the beginning, once the last
item has been reached.
Digital Data Area
The digital data area of the Bedside tab displays a maximum of eight digital parameters.
Parameters experiencing an alarm condition will display a different background color
indicating the severity of the alarm. For additional information, see ‘‘Alarm Behavior’’ on
page 3-1. If ST is enabled on the ECG waveform, the digital data section of the Bedside tab
includes an ST lead label that specifies the lead from which the displayed data is acquired.
7.1.3 Sidebar Buttons
The sidebar buttons are used to expand the functionality of this tab.
7.1.3.1 View More
Select the View More sidebar button to scroll additional waveforms and/or digital data
into the bottom position on the screen.
7.1.3.2 Mark Event
Select the Mark Event sidebar button to save the current data to the Event database as a
Central User Marked Event. For example, the Mark Event sidebar button can be useful for
noting such things as the time when a drug was administered.
7.1.3.3 Freeze
Select the Freeze sidebar button to freeze all of the waveforms and parameter data that are
currently displayed on the Bedside tab. The Bedside tab remains frozen until the Freeze
sidebar button is selected again, even if another patient is selected. Freezing the Bedside
tab does not affect the waveform or any real time alarms in the upper half of the normal
screen.
When the Freeze sidebar button is selected, the Calipers sidebar button is enabled. This
replaces the Mark Event sidebar button.