Panorama™ Operating Instructions 0070-00-0575-50 6 - 17
Patient Alarm Functions Arrhythmias
The Trigeminy alarm priority settings include 2, 3 and OFF and behaves as follows:
If the Trigeminy alarm priority is OFF, the Panorama Central Station does not produce
any audio alarm indicators for the Trigeminy alarm condition.
If the Trigeminy alarm priority is set to 2, the Panorama Central Station produces an
Alarm Priority 2 sound once the condition is identified.
If the Trigeminy alarm priority is set to 3, the Panorama Central Station produces an
Alarm Priority 3 sound once the condition is identified.
The Trigeminy alarm is associated with a yellow Trigeminy alarm text message in the
Patient Status line.
For additional information, refer to ‘‘Alarm Behavior’’ on page 3-1.
Triplet Alarm
The Triplet alarm is activated when the Panorama Central Station identifies three
consecutive PVCs between normal beats.
NOTE: A Triplet alarm is not a latched alarm and can be
acknowledged at anytime.
NOTE: The Triplet alarm is not available for the Neonate patient
size.
The Triplet alarm priority settings include 2, 3 and OFF and behaves as follows:
If the Triplet alarm priority is OFF, the Panorama Central Station does not produce any
audio alarm indicators for the Triplet alarm condition.
If the Triplet alarm priority is set to 2, the Panorama Central Station produces an Alarm
Priority 2 sound once the condition is identified.
If the Triplet alarm priority is set to 3, the Panorama Central Station produces an Alarm
Priority 3 sound once the condition is identified.
The Triplet alarm is associated with a yellow Triplet alarm text message in the Patient
Status line.
For additional information, refer to ‘‘Alarm Behavior’’ on page 3-1.
V-Rhythm Alarm
The V-Rhythm alarm is activated when the Panorama Central Station a number of
consecutive PVCs at a rate less than the V-Tach threshold.
NOTE: A V-Rhythm alarm is not a latched alarm and can be
acknowledged at anytime.
NOTE: The V-Rhythm alarm is not available for the Neonate patient
size.