User's Manual

Arrhythmias Patient Alarm Functions
6 - 16 0070-00-0575-50 Panorama™ Operating Instructions
NOTE: The Missed Beat alarm is not available for the Neonate
patient size.
The Missed Beat alarm priority settings include 2, 3 and OFF and behaves as follows:
• If the Missed Beat alarm priority is OFF, the Panorama Central Station does not produce
any audio alarm indicators for the Missed Beat alarm condition.
• If the Missed Beat alarm priority is set to 2, the Panorama Central Station produces an
Alarm Priority 2 sound once the condition is identified.
• If the Missed Beat alarm priority is set to 3, the Panorama Central Station produces an
Alarm Priority 3 sound once the condition is identified.
• The Missed Beat alarm is associated with a yellow Missed Beat alarm text message in
the Patient Status line.
For additional information, refer to ‘‘Alarm Behavior’’ on page 3-1.
Run Alarm
The Run alarm is activated when the Panorama system identifies a user-defined number of
consecutive PVCs at a heart rate of between 100 to 150 bpm.
NOTE: A Run alarm is not a latched alarm and can be
acknowledged at anytime.
NOTE: The Run alarm is not available for the Neonate patient size.
The Run alarm priority settings include 2, 3 and OFF and behaves as follows:
• If the Run alarm priority is OFF, the Panorama Central Station does not produce any
audio alarm indicators for the Run alarm condition.
• If the Run alarm priority is set to 2, the Panorama Central Station produces an Alarm
Priority 2 sound once the condition is identified.
• If the Run alarm priority is set to 3, the Panorama Central Station produces an Alarm
Priority 3 sound once the condition is identified.
• The Run alarm is associated with a yellow Run alarm text message in the Patient Status
line.
For additional information, refer to ‘‘Alarm Behavior’’ on page 3-1.
Trigeminy Alarm
The Trigeminy alarm is activated when the Panorama Central Station identifies two or more
cycles of one PVC followed by two normal beats.
NOTE: A Run alarm is not a latched alarm and can be
acknowledged at anytime.
NOTE: The Run alarm is not available for the Neonate patient size.