User's Manual

Alarm Responses Tab (Patient) Patient Alarm Functions
6 - 12 0070-00-0575-50 Panorama™ Operating Instructions
3. Select the Alarm Delay time period for the selected patient’s alarm. Alarm Delay is the
period of time that the Panorama Central Station delays the notification of an alarm.
Alarm Delay options include None, 1 second, 2 seconds, 3 seconds, 4 seconds, 5
seconds, 6 seconds, 7 seconds and 8 seconds.
NOTE: The Alarm Delay setting does not affect the alarm
notification for ST, Apnea or Arrhythmia.
4. Select the Done or the Cancel button.
•The Done button accepts the configured alarm delay settings and returns to the
Alarm Responses tab.
•The Cancel button discards the configured alarm delay settings and returns to the
Alarm Responses tab.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, close
all tabs, message boxes and dialog boxes.
6.3.4 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Patient Alarm Responses tab.
MESSAGE/ISSUE * REASON SOLUTION
No patient Selected! This message appears in the
demographic field when either
no patient tile or an empty
patient tile has been selected.
Select the OK button to close
the dialog box. Select an
active patient tile.
Communication Lost Communication between the
Panorama Central Station and
the monitoring device for the
selected patient tile has been
interrupted.
Check the connection between
the Panorama Central Station
and the monitoring device. For
example is the monitoring
device out of range?
All of the sidebar buttons
(except the Normal Screen
button) are disabled.
An empty patient tile is
selected.
Select a valid patient tile by
touching the digital or
waveform area of the patient
tile.
All Alarms Suspended The Suspend All Alarms
function is enabled. All alarm
processing for the selected tile
has been suspended.
Resume alarm processing by
pressing the Suspend All
Alarms button and disabling
the Suspend All Alarms
function.
Non-Lethal Alarms
Suspended
The Suspend Non-lethal
function is enabled. Alarm
processing for numeric,
technical and non-lethal alarms
has been suspended for the
selected tile.
Resume alarm processing by
pressing the Suspend Non-
lethal button and disabling
the Suspend Non-lethal
function.
* Messages are shown in all bold text.