Panorama™ Operating Instructions 0070-00-0575-50 6 - 11
Patient Alarm Functions Alarm Responses Tab (Patient)
To resume the monitoring of the alarms at the Panorama Central Station select the Suspend
All Alarms sidebar button.
• When the Suspend All Alarms sidebar button is disabled, it is blue with white text.
Alarms are now re-enabled for the patient.
Suspend Non-lethal
Select the Suspend Non-lethal sidebar button to suspend all numeric, technical and non-
lethal arrhythmia alarms for the selected patient tile. Lethal arrhythmia alarms continue to be
monitored at the Panorama Central Station when non-lethal arrhythmia alarms are
suspended.
WARNING: The Panorama Central Station will not monitor any non-
lethal arrhythmia, numeric or technical alarms for the
selected patient tile when the Suspend All Non-lethal Alarms
button is enabled.
To suspend all non-lethal alarms select the Suspend Non-lethal sidebar button.
• When the Suspend Non-lethal sidebar button is enabled, the button is white with
black text.
•The Non-Lethal Alarms Suspended message displays in yellow text in the Patient
Status line of the patient tile.
To resume monitoring of non-lethal alarms, select the Suspend Non-lethal sidebar button.
• When the Suspend Non-lethal sidebar button is disabled, it is blue with white text.
Alarms are now re-enabled for the patient.
Alarm Setup (Optional)
Select the Alarm Setup sidebar button to configure the ST Alarm Delay and the Alarm
Delay for the selected patient tile. The Alarm Setup sidebar button only displays when the
Bedside Alarm Tracking option is OFF in the System Alarms tab.
1. Select the Alarm Setup button. The Alarm Setup dialog box (FIGURE 6-4) is
displayed.
FIGURE 6-4 Alarm Setup Dialog Box
2. Select the ST Alarm Delay time period for the selected patient’s alarm. ST Alarm
Delay is the period of time that the Panorama Central Station delays the notification of
an ST alarm. ST Alarm Delay options include 30 seconds, 45 seconds, 1 minute, 90
seconds, 2 minutes and 3 minutes.