Alarm Responses Tab (Patient) Patient Alarm Functions
6 - 10 0070-00-0575-50 Panorama™ Operating Instructions
3. Set the alarm priority for the alarm parameter. To set an alarm priority, select the
Priority button until the alarm priority setting is shown. Alarm priority is associated with
the severity of an alarm condition. It determines the audio and visual indicators that
occur during an alarm condition.
Alarm priority settings may include Off, 1, 2 and 3. Each alarm has predetermined
alarm priority settings.
•Priority1 are the most severe alarms (lethal alarms).
•Priority 2 are alarms of medium severity (most numeric alarms).
•Priority 3 are the least severe alarms (technical alarms).
•Priority OFF disables audio feedback for the selected alarm at the Panorama Central
Station.
For additional information on alarm behavior, refer to ‘‘Alarm Behavior’’ on page 3-1.
4. The Print On Alarm button sets the automatic printing of alarm data during an alarm
condition.
•The ON option automatically prints the alarm events when the alarm condition occurs.
•The OFF option suspends the printing of alarm events during an alarm condition.
5. The Save To Event button sets the automatic saving of alarm data in the Event
database when an alarm condition occurs.
•The ON option automatically saves alarm data during an alarm condition.
•The OFF option does not save alarm data during an alarm condition.
6.3.3 Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Restore Previous Settings
Select the Restore Previous Settings sidebar button to return patient display settings to
those in use when the tab was opened.
Suspend All Alarms (Optional)
Select the Suspend All Alarms sidebar button to suspend all numeric, lethal arrhythmia,
non-lethal arrhythmia and technical alarms at the Panorama Central Station for the selected
patient tile. The Suspend All Alarms sidebar button is only shown if the Suspend Level
1 option is ON in the System Alarms tab.
WARNING: The Panorama Central Station will not monitor any non-
lethal arrhythmia, numeric or technical alarms for the
selected patient tile when the Suspend All Non-lethal Alarms
button is enabled.
To suspend all alarms at the Panorama Central Station, select the Suspend All Alarms
sidebar button.
When the Suspend All Alarms sidebar button is enabled, the button is white with
black text.
•The All Alarms Suspended message displays in red text in the Patient Status line of
the patient tile.