User's Manual

Alarm Limits Tab (Patient) Patient Alarm Functions
6 - 8 0070-00-0575-50 Panoramaâ„¢ Operating Instructions
6.2.4 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Patient Alarm Limits tab.
MESSAGE/ISSUE * REASON SOLUTION
No patient Selected! This message appears in the
demographic field when either
no patient tile or an empty
patient tile has been selected.
Select an active patient tile.
Communication Lost Communication between the
Panorama Central Station and
the monitoring device for the
selected patient tile has been
interrupted.
Check the connection between
the Panorama Central Station
and the monitoring device. For
example is the monitoring
device out of range?
All of the sidebar buttons
(except the Normal Screen
button) are disabled.
An empty patient tile is
selected.
Select a valid patient tile by
touching the digital or
waveform area of the patient
tile.
All Alarms Suspended The Suspend All Alarms
function is enabled. All
numeric, lethal arrhythmia,
non-lethal arrhythmia and
technical alarm processing for
the selected tile has been
suspended.
Resume alarm processing by
pressing the Suspend All
Alarms button and disabling
the Suspend All Alarms
function.
Non-Lethal Alarms
Suspended
The Suspend Non-lethal
function is enabled. Alarm
processing for numeric,
technical and non-lethal alarms
has been suspended for the
selected tile.
Resume alarm processing by
pressing the Suspend Non-
lethal button and disabling
the Suspend Non-lethal
function.
* Messages are shown in all bold text.