User's Manual

Wave Gain Tab (Patient) Patient Setup Functions
5 - 38 0070-00-0575-50 Panorama™ Operating Instructions
4. Select the Resp Gain button until the desired wave gain is displayed. The settings
available for the resp gain wave gain include 1, 2, 3, 4 and 5.
5. Select the CO2 (%, kPa, mmHG) button until the desired wave gain is displayed.
• The settings available for the CO
2
% wave gain include 5.0, 7.5 and 10.0 %.
• The settings available for the CO
2
kPa wave gain include 5.0, 7.5 and 10.0 kPa.
• The settings available for the CO
2
mmHg wave gain include 40, 60 and 100 mmHg.
6. Select the O2 % button until the desired wave gain is displayed. The settings available
for the O
2
wave gain include 18 - 30, 18 - 60 and 18 - 100 %.
7. Select the AGENT % button until the desired wave gain is displayed. The settings
available for the agent wave gains include 0 - 1.0, 0 - 2.5, 0 - 5.0, 0 - 10.0,
0 - 15.0 and 0 - 20.0 %.
8. Select the Done button to accept the wave gain settings shown in the Wave Gain tab.
A patient’s wave gain settings will not be changed unless the Done button is selected.
5.6.3 Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Restore Previous Settings
Select the Restore Previous Settings sidebar button to return patient display settings to
those in use when the tab was opened.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, close
all tabs, message boxes and dialog boxes.
5.6.4 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Patient Wave Gain tab.
MESSAGE/ISSUE * REASON SOLUTION
No patient Selected! This message appears in the
demographic field when either
a patient tile has not been
selected or an empty tile was
selected.
Select an active patient tile.
The patient’s wave gain
settings did not change
The Done button was not
selected after the patient’s
wave gain settings were
changed.
Adjust the patient’s wave gain
settings and select the Done
button when the settings are
completed.
* Messages are shown in all bold text.