User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 5 - 37
Patient Setup Functions Wave Gain Tab (Patient)
5.6 Wave Gain Tab (Patient)
The Wave Gain tab (FIGURE 5-24) is used to adjust the wave gain settings for the selected
patient’s available waveforms. The initial wave gain settings for a patient are the system
default settings at the time the patient is admitted.
NOTE: The layout and functionality of the Wave Gain tab changes
based on the type of monitoring device assigned to the
selected patient tile. The fields discussed in this section may
not apply to all users.
This section includes:
Accessing the Wave Gain Tab (Patient)
Adjusting a Patient’s Wave Gain Settings
Sidebar Buttons
Troubleshooting
5.6.1 Accessing the Wave Gain Tab (Patient)
1. From the menu bar, select the Patient Setup menu button. The tabs associated with
patient setup display.
2. Select the Wave Gain tab. The Wave Gain tab is displayed.
FIGURE 5-24 Wave Gain Tab
5.6.2 Adjusting a Patient’s Wave Gain Settings
This section outlines entering and modifying a patient’s wave gain settings via the Wave
Gain tab.
1. Choose a patient tile by selecting its digital data or waveform area.
2. In the Wave Gain tab, select the ECG mm/mV button until the desired wave gain
displays. The settings available for the ECG wave gain include 2.5, 5.0, 10.0, 20.0
and 30.0 mm/mV.
3. Select the IBP mmHg button until the desired wave gain displays. The settings
available for the IBP wave gain include 0 - 20, 0 - 40, 0 - 80, 60 - 140, 0 - 160, 0 -
320 and (-10) - 10 mmHg.
NOTE: If the IBP is labeled at the bedside monitor, the same label
displays on the wave gain button.