User's Manual
Monitor Application Getting Started
4 - 8 0070-10-0678-01 NetGuard™ Operating Instructions
•The icon indicates that the alarm tone was muted.
•The Elapsed Time indicates the amount time that has passed since the alarm was
called. HH:MM:SS is the time format displayed.
•The ECG button indicates that there is an ECG waveform associated with the selected
alarm tile. Selecting the ECG button opens the ECG Waveform window (shown in
FIGURE 4-8). The ECG Waveform window contains 20 seconds of waveform data. The
blue line in the window indicates when the server received the alarm notification.
FIGURE 4-8 ECG Waveform Window
Primary Alarm Window Buttons
The Primary Alarm window will display two (2) or three (3) buttons depending on the alarm
type. System level alarms will display a MUTE and OVERRIDE button. Patient level alarms
will display a MUTE, OVERRIDE, and DEFER button. The buttons are described in
Table 4-1:
4.3.2.3 Deferred Alarm Application Bar
The Deferred Alarm Application Bar (shown in FIGURE 4-9) displays active technical and
deferred Patient level alarms. If there are no active technical or deferred Patient level alarms,
the Deferred Alarm Application Bar will not be displayed.
NOTE: Deferred alarms are terminal specific. Alarms deferred at
one terminal are not deferred at other terminals.
TABLE 4-1
BUTTON DESCRIPTION
MUTE Silences an active alarm’s audio alerting tone for
30 seconds.Once MUTE is selected, the countdown time is
displayed inside the button.
OVERRIDE After authentication, closes the currently active alarm window.
Selecting this button is equivalent to acknowledging the alarm at
the device.
DEFER Moves the alarm to the Deferred Application Bar.
NOTE: The DEFER button is only displayed for
Patient level alarms. Due to their severity,
System level alarms cannot be deferred.
Blue Line