User's Manual

NetGuard™ Operating Instructions 0070-10-0678-01 4 - 5
Getting Started Monitor Application
4.3.2 Alarm Alert View
The Alarm Alert view displays physiological and technical alarm events. This view is only
seen when an alarm is triggered or when the mute duration for an alarm has expired. The
Alarm Alert view consists of the Primary Alarm window and the Deferred Alarm Application
Bar.
FIGURE 4-4 Alarm Alert View - System Alarms (multiple alarm tiles)
4.3.2.1 Primary Alarm Window
The Primary Alarm window displays System level and Patient level alarm events.
NOTE: Under certain circumstances, the Primary Alarm Window
may also display technical events.
In the event of a System level alarm, a window will pop-up covering all open applications,
and the Microsoft
®
Windows taskbar and navigation buttons. The system displays System
level alarms in red, and displays a icon beside the heart rate to indicate that the alarm
is high priority. The Primary Alarm window cannot be closed until the alarm is either
acknowledged or overridden.
The Primary Alarm window will also pop-up for Patient level alarms but will not entirely cover
the screen. The standard Microsoft
®
Windows taskbar and navigation buttons will still be
available. Patient level alarms may be red or yellow depending on system configuration
settings. Patient level alarms that are configured to display in red will display an icon.
Patient level alarms configured to display in yellow will display an icon indicating a
lower alarm priority. Patient level alarms may be deferred, overridden, or muted.
Primary Alarm Window
Deferred Alarm
Application Bar
Frame Region
Alarm Tile List