Instruction manual

If the span adjust value is within the accuracy requirements of the weighing system,
the material test was successful and normal operation can be resumed.
If the span adjust value is not acceptable, repeat the material test to verify
repeatability. If the result of the second material test differs considerably, consult
Milltronics or its agent.
If the span adjust values are significant and repeatable, perform a manual span adjust:
refer to maneuvering for
parameter selection
and value change
Verify the results of the span adjust by material test or return to normal operation.
DESIGN CHANGES
Where parameters have been changed with a resultant impact on the calibration or do
not take effect until a recalibration is done, the parameter warning icon is displayed. In
order to clear the icon, perform a zero and span recalibration after the reprogramming
session is complete.
If significant changes have been made, an initial zero (P377) and/or initial span
(P388) may be required.
98/03/10
program mode,
P019
enter % span adjust,
e.g. 0.05
jumps to P017 adjusted
test load,
e.g. 97.95
if % span adjust is
negative, be sure
to enter as
negative value
RECALIBRATION
PL-531 37