Instruction manual

Place the BW100 into the program mode at P291.
Remove the test weight, insure that the BW100 SW2 is in ‘close’ position and that P1
is left as set.
Performing a balance procedure requires a subsequent
zero and span recalibration.
‘A’
‘B’
test weight
test weight on
load cell ‘A’ side
enter and auto advance to P295
‘close’
SW2
enter and auto advance to P293,
leave SW2 open
P1
adjust until
## = 0
± 5
‘A’
‘B’
test weight
test weight on
load cell ‘B’ side
enter and auto advance to P292,
open SW2
‘close’
SW2
select P291 ensure SW2
is closed
SW2
‘open’
enter and auto advance to P294,
close SW2
SW2
‘open’
START UP
PL-531 27