User manual

70
Description, Continued
Set Points
Diagram 1 Diagram 2
Item Description
Pretreatment Change set point for controlling the
frequency of the message REPLACE
EXTERNAL PRE-TREATMENT.
Strainer Frequency Change set points for controlling the
frequency of the message EXAMINE
INLET STRAINER.
Tap Feed Cond Change set point controlling the message
TAP FEED CONDUCTIVITY > SP.
RO Rejection Change set point controlling the message
RO REJECTION < SP.
Permeate Cond Change set point controlling the message
PERMEATE C > SP.
Elix
Product Res
Change set point controlling the message
ELIX PRODUCT R < SP.
Tank Refill Change set point controlling the tank level
where the Milli-Q System starts to refill
the tank.
Milli-Q Product Res Change set point controlling the message
MILLI-Q RES < SP, REPLACE
QUANTUM.
Milli-Q Product TOC Change set point controlling the message
MILLI-Q TOC > SP.
Millipak Change set point controlling the message
REPLACE Q-POD (or E-POD) PAK X IN
YY DAYS (where 1 d X d 3 and
1 d YY d 14).
BioPak See above.
POD Pak A, POD Pak B,
POD Pak C
See above.