User manual

61
Cleaning the A10 TOC Monitor
Introduction
The A10 TOC Monitor is cleaned whenever a new Quantum Cartridge is
installed.
Procedure
Follow the steps below to clean the A10 TOC Monitor.
Step Action Diagram
1 Start in STANDBY Mode.
2
x Select <Menu>.
x Press
.
3
x Select <Sanitise/Clean>.
x Press
.
4
x Select <A10 Cleaning>.
x Press
.
Continued on next page