User manual

58
Calibrating the Flowrate
Introduction
x The Milli-Q Water flowrate should be calibrated when the Milli-Q System
is installed.
x This calibration should be done with Q-POD Unit 1 in case there are
multiple Q-POD Units.
x A 1 Litre graduated cylinder is needed.
Procedure
Follow the steps below to perform a Flow Calibration.
Step Action Diagram
1 Go to STANDBY Mode.
2
x Select <Menu>.
x Press
.
3
x Enter the <Manager Menu>.
x See the Software Chapter to
learn how to enter the
Manager Menu.
4
x Select <Setup>.
x Press
.
Continued on next page