Alarms generated during installation
Overview
x During the installation of a Milli-Q System, certain Alarm messages are
generated.
x This occurs because:
the Reservoir is empty,
there is air in the tubings and in the Progard Pack,
the Progard Pack is not installed,
the Quantum Cartridge is not installed, and
an automatic flush of the Progard Pack is manually cancelled.
x These alarms are explained here. The ways to cancel them are explained
also. For more information about Alarm messages, see the chapter titled
‘Alarms’.
x It is perfectly normal to see alarms during installation.
x The Milli-Q System is designed to use various sensors to alert you of
problems during normal operation of the system. This insures optimal
water quality.
x During installation, these sensors are active. As a result, it is possible to
have alarms generated. In order to advance during the installation, these
alarms should be cancelled.
TANK EMPTY
message
x This alarm occurs because the Reservoir is empty during most of the
installation.
x This alarm goes away when the Reservoir is partially full.
x To cancel the text display of this alarm message, follow the instructions on
the LCD.
PROGARD
PACK OUT
message
x This alarm occurs because the Progard Pack is not installed.
x This alarm goes away when the Progard Pack is detected by the Milli-Q
System.
x To cancel the text display of this alarm message, follow the instructions on
the LCD.
QUANTUM
CARTRIDGE
OUT message
x This alarm occurs because the Quantum Cartridge is not installed.
x This alarm goes away when the Quantum Cartridge is detected by the
Milli-Q System.
x To cancel the text display of this alarm message, follow the instructions on
the LCD.