147
Summary of Alert messages
Alert messages
LCD message What it means
CALIBRATION VISIT OVERDUE
XX DAYS
x The Milli-Q System has
determined that a Calibration Visit
is overdue.
x Contact Millipore.
CHECK ASM UV LAMP
x The ASM UV Lamp is not turning
on.
x Contact Millipore.
CHECK UV 185 NM LAMP
x The UV 185 nm Lamp is not
turning on.
x Contact Millipore.
CHECK UV 254 NM LAMP
x The UV 254 nm Lamp is not
turning on.
x Contact Millipore.
EDI LOW INTENSITY
x The Milli-Q System has
determined that the electrical
intensity used by the EDI Module
is below specification.
x Contact Millipore.
EXAMINE INLET STRAINER
x The Milli-Q System has
determined that it is time to clean
the Inlet Strainer.
x Clean the Inlet Strainer and reset
the message.
NEXT CALIBRATION VISIT IN
XX DAYS
x The Milli-Q System is prompting
you that a Calibration Visit should
be scheduled.
x Contact Millipore.
NEXT QUALIFICATION VISIT
IN XX DAYS
x The Milli-Q System is prompting
you that a Qualification Visit
should be scheduled.
x Contact Millipore.
NEXT SERVICE VISIT IN XX
DAYS
x The Milli-Q System is prompting
you that a Service Visit should be
scheduled.
x Contact Millipore.
Continued on next page