User manual
136
Summary of Alarm messages, Continued
Alarm
messages
(continued)
LCD message What it means
LOW FEED WATER PRESSURE
x Check Feedwater pressure and
rectify.
x Go to STANDBY Mode and go to
READY Mode to release any
trapped air in the Milli-Q System.
x Contact Millipore if problem
persists.
MILLI-Q RES < SP, REPLACE
QUANTUM
x The Milli-Q Water resistivity is <
set point.
x Dispense water to eliminate any
trapped air in the Milli-Q System.
x Replace the Quantum Cartridge.
MILLI-Q RES > MAX
x The Milli-Q Water resistivity is
out of measurement range.
x Contact Millipore.
MILLI-Q T < MIN
x The Milli-Q Water temperature is
out of measurement range.
x Contact Millipore.
MILLI-Q T > MAX
x The Milli-Q Water temperature is
out of measurement range.
x Contact Millipore.
MILLI-Q TOC > SP
x The TOC is > set point.
x Contact Millipore.
NO FLOW MEASURE
x The Flowmetre has measured <
0.2 Lpm during Volumetric
Dispensing mode.
x Push the Q-POD Unit Plunger all
the way down and release.
PERMEATE C < MIN
x The Permeate conductivity is out
of measurement range.
x Contact Millipore.
PERMEATE C > MAX
x The Permeate conductivity is out
of measurement range.
x Contact Millipore.
PERMEATE C > SP
x The Permeate conductivity is > set
point.
x Contact Millipore.
Continued on next page