Summary of Alarm messages, Continued 
Alarm
messages 
(continued)
LCD message  What it means 
LOW FEED WATER PRESSURE 
x Check Feedwater pressure and 
rectify.
x Go to STANDBY Mode and go to 
READY Mode to release any 
trapped air in the Milli-Q System. 
x Contact Millipore if problem 
persists. 
MILLI-Q RES < SP, REPLACE 
QUANTUM 
x The Milli-Q Water resistivity is < 
set point. 
x Dispense water to eliminate any 
trapped air in the Milli-Q System. 
x Replace the Quantum Cartridge. 
MILLI-Q RES > MAX 
x The Milli-Q Water resistivity is 
out of measurement range. 
x Contact Millipore. 
MILLI-Q T < MIN 
x The Milli-Q Water temperature is 
out of measurement range. 
x Contact Millipore. 
MILLI-Q T > MAX 
x The Milli-Q Water temperature is 
out of measurement range. 
x Contact Millipore. 
MILLI-Q TOC > SP 
x The TOC is > set point. 
x Contact Millipore. 
NO FLOW MEASURE 
x The Flowmetre has measured < 
0.2 Lpm during Volumetric 
Dispensing mode. 
x Push the Q-POD Unit Plunger all 
the way down and release. 
PERMEATE C < MIN 
x The Permeate conductivity is out 
of measurement range. 
x Contact Millipore. 
PERMEATE C > MAX 
x The Permeate conductivity is out 
of measurement range. 
x Contact Millipore. 
PERMEATE C > SP 
x The Permeate conductivity is > set 
point.
x Contact Millipore. 
Continued on next page 