User manual
135
Summary of Alarm messages, Continued
Alarm
messages
(continued)
LCD message What it means
CLEANING CANCELLED
x A cleaning mode was cancelled
and was not fully completed.
x Go to STANDBY Mode and then
go to READY Mode.
x The Milli-Q System will go into a
15 minute FLUSH Mode. The
system will then automatically fill
the Reservoir.
ELIX PRODUCT R < SP
x The resistivity of the water filling
the Reservoir is < set point.
x Contact Millipore.
ELIX PRODUCT R > MAX
x The resistivity of the water filling
the Reservoir is out of
measurement range.
x Contact Millipore.
ELIX PRODUCT T < MIN
x The temperature of the water
filling the Reservoir is out of
measurement range.
x Contact Millipore.
ELIX PRODUCT T > MAX
x The temperature of the water
filling the Reservoir is out of
measurement range.
x Contact Millipore.
FLOW AUTO STOP
x The Milli-Q System has
automatically stopped dispensing
water. The POD FLOW STOP
timer has reached 0 minutes.
x Push the Q-POD Unit Plunger all
the way down and release.
x This resets the dispenser timer and
makes the Q-POD Unit available
for dispensing.
INCORRECT PROGARD PACK
x The Milli-Q System does not
recognise the type of Progard Pack
being installed.
x Contact Millipore.
INCORRECT QUANTUM
CARTRIDGE
x The Milli-Q System does not
recognise the type of Quantum
Cartridge being installed.
x Contact Millipore.
Continued on next page