User manual

133
Summary of Alarm messages
Alarm
messages
LCD message What it means
A10 ERROR 0
x A10 PCB E
2
Prom defective.
x Power off, power on the Milli-Q
System.
x Dispense water for several
minutes.
x If the message continues, then
contact Millipore.
A10 ERROR 1
x A10 PCB A/D converter
defective.
x Power off, power on the Milli-Q
System.
x Dispense water for several
minutes.
x If the message continues, then
contact Millipore.
A10 ERROR 2
x A10 Thermistor defective.
x Power off, power on the Milli-Q
System.
x Dispense water for several
minutes.
x If the message continues, then
contact Millipore.
A10 ERROR 3
x Problem occurred with
temperature compensation.
x Power off, power on the Milli-Q
System.
x Dispense water for several
minutes.
x If the message continues, then
contact Millipore.
A10 ERROR 4
x The water entering the A10 is <
4°C.
x If the problem can not be resolved,
then contact Millipore.
A10 ERROR 5
x The water entering the A10 is >
41°C.
x If the problem can not be resolved,
then contact Millipore.
Continued on next page