User manual

MAINTENANCE
- 61 -
H
OW TO RESET THE UV LAMP TIMER
4. Press the Main
and “+“ Buttons
together to enter
the menu. The
Display will
show C01.
5. Press the Main
Button 4 times.
The Display will
show C05 and
“0” days left on
the UV Timer.
6. Press the “+”
and “-“ Buttons
together. This
will reset the
UV Timer to
500 (days). The
Display will exit
the menu.
The UV Timer has been reset.