User manual
PRE INSTALLATION
- 7 -
E
NVIRONMENTAL REQUIREMENTS
Indoor Use Only
Storage Temperature
5 °C < T < 40 °C
Operating Temperature
5 °C < T < 40 °C
Relative Humidity
Should not exceed 80% for temperatures below 31 °C Should
stay within 50% to 80% between 31 °C and 40 °C.
Altitude < 3000 metres
Installation Category II
Pollution Degree 2
3-2 OPTIONAL EQUIPMENT YOU MAY NEED
WALL MOUNTING BRACKET
The Millipore Catalogue Number for the Wall Mounting Bracket is WMBSMT001. The mounting hardware
for attaching the bracket to a wall is not included and must be supplied.
E
XTERNAL TANK CONNECTOR
The Millipore Catalogue Number for a Tank Connector Kit is TANKPECKT. An external tank holding up
to 30 Litres can be used instead of the internal 6 Litre tank. The Tank Connector Kit is used to connect the
30 L tank to the system. The mounting hardware for connecting the external tank to the system is not
included and must be supplied. The Millipore Catalogue Number for the 30 L PE Reservoir is TANKPE030.