User's Manual

Table Of Contents
3-8 Millennial Net
Running MeshScape Network Monitor
Configuring the Sample Interval of a Single Node
To configure the time interval between data packets transmitted by a node (see Figure 3-12):
1. Double-click on the desired device from the list of discovered sensor nodes. The Device
window is opened, displaying the device’s current configuration.
2. Using the Sampling Interval panel, enter the interval sampling rate as a multiple of 100
milliseconds. For example, to configure a sampling interval rate of 10 seconds, enter a
value of 100.
minimum value: 1 (0.1 sec)
maximum value: 65535 (109 minutes)
3. Select Update. The device’s configuration is updated.
4. Select X to exit the Device window.
Figure 3-12. Configuring sample interval of single node
Note:
After you click Update, the MeshScape Network Monitor will display the ‘Refresh’
state for the device. Changes to the device configuration will show up after one
sampling interval. You can use the Watch window to track configuration changes.
Double-click on device
1
Configure interval time
2
Select Update
3
Select X
4