User Manual
2
software and firewall.
Before use the Miko Drip Management system, please confirm the capacity of device
licenses and due date.
You should read the instructions before using the equipment.
When the equipment shows abnormal warnings or behaviors, please reboot the default
settings. If reboot the default settings cannot make the equipment work, please stop use
equipment and contact the original manufacturer or distributor to repair or update. Please
do not disassemble or replace parts by yourself; otherwise, it may cause electric shock or
accidental injury.
Before using the equipment, please confirm whether the batteries are placed correctly.
Do not dismantle or modify any electronic parts or wires. If the device is in need for
repair or maintenance, contact the original manufacturer or distributor.
Depending on the frequency of use, the operator should regularly contact the original
manufacturer or distributor for machine calibration to ensure the accuracy of the weight
measured.
Please do weight calibration regularly.
When the drip monitoring management system is crashed, the alarm mechanism only
shows by the light and sound the equipment.
Be sure the pairing of equipment and bed number is correct.
Set the correct weight of infusion set to avoid false alarms.
Please remind patients don’t adjust the equipment settings to avoid false alarms.
Please use the correct software version when using the equipment.
Do not uninstall the software arbitrarily to avoid failure of monitoring.
In order to ensure the accuracy of the weight monitoring, please hang the IV drip on the
equipment. Be sure that the hanging method is correct.
Do not hang other heavy objects with IV Drip on the equipment at the same time.
Use the equipment in the acceptable environment temperature.
Do not hang heavy objects exceeding 2500g, because it may cause the weight sensing
mechanism fail.
When you see battery warning on the system, please replace the batteries immediately.
The operator should not touch the patient and the equipment at the same time, especially
when changing batteries.
Dropping or improperly usage may cause damage of equipment.Please do not take a
hammer or other items to hammer blow, or throw device and trampling, falls, and fall
causing a strong impact, heavy blow, etc.
Notice : The changes or modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate the equipment.
IMPORTANT NOTE: To comply with the FCC RF exposure compliance requirements, no
change to the antenna or the device is permitted. Any change to the antenna or the