Technical data
If a Disinfection programme is running, the following
message is displayed:
Fault
Container empty
O
OK
H
The programme will be cancelled.
^
Press short cut button D to delete the message in the
display, and refill the container.
^
Start the programme again.
,
If the container is empty for a disinfection programme,
you have to assume that no disinfection has taken place.
Start the programme again.
Detergent
33