Technical data
9. Installation & activation/deactivation of the water intake connections
All plumbing work must be performed by licensed professionals and adhere to all applicable Local, State, and Federal Codes.
The Miele G 7881 Dental Disinfector is equipped with both a DI Water intake (marked “H2O PUR”) hose and a
Cold Tap Water intake (marked with a blue sticker) hose. Leading instrument manufacturers recommend a DI
water rinse as the final processing step prior to sterilization. Rinsing with DI water ensures the elimination of
detergent residues and naturally occurring minerals which could lead to instrument stains and corrosion. However,
if water quality is not an issue, the machine can be installed using the Cold Tap Water intake only (see below).
1. Turn off the machine with the "On/Off" button.
2. Turn the program selector to the "Stop" position.
3. Press and hold the "Display” and "Start” buttons at the same time.
4. Simultaneously turn the machine on with the "On/Off" button.
The current program status "P…" will appear in the display and the "Fill/Drain” indicator will illuminate.
5. Press the "Drying” button once. "E01" (programming level 1) will appear in the display.
6. Turn the program selector one switch position counter-clockwise (11 o’clock position).
If machine is installed with both the DI Water & Cold Tap Water intakes:
7. Press the "Display” button as many times as necessary until the display shows the number “110.”
Once the number "110" has been reached the G 7881 will activate the DI Intake Valve for the final rinse.
8. Press the "Start” button. "SP" will appear in the display.
9. Press the "Start” button again.
OR:
If machine is installed with only the Cold Tap Water intake:
7. Press the "Display" 4 button as many times as necessary until the display shows the number “111.”
Once the number "111" has been reached the G 7881 will deactivate the DI Intake Valve.
8. Press the "Start” button. "SP" will appear in the display.
9. Press the "Start" button again.
10. Adjust the feet to the proper height/Install machine on solid non-carpeted surface
●
Undercounter
Note: Removal of lid not necessary for all undercounter installations.
Open the door, remove fixing screws on the left and right sides using a Phillips head screwdriver.
Pull the machine lid approx. 5mm forwards, lift it upwards, and remove.
Adjust feet to proper height.
Slide machine into place, making sure machine is level in all directions.
Secure machine to the front edge of the worktop using screws to the left and right of the front trim.
Do not use silicone sealant in gaps between the machine & neighboring units.
Secure machine to countertop, making certain to not pinch hoses.
●
Freestanding
Use Mounting Kit (Item #00253531)
11. Install the Steam Deflector if applicable
Note: A Steam Deflector protects the countertop and comes packaged with the unit.
Miele
●
9 Independence Way
●
Princeton, NJ 08540
●
609-419-9898
●
800-991-9380
●
Fax 609-419-4241
●
www.dentaldisinfector.com
4
The inlet hose must not be shortened or damaged in any way.
Acceptable water pressure (flow rate) for both intake valves: 10-147 psi each
Recommended water pressure (flow rate) for both intake valves: 25-60 psi each
●
DI Water Intake: ¾"(19 mm) GHT female hose thread, connected to external DI water supply source
●
Cold Tap Water Intake: ¾"(19 mm) GHT female hose thread, connected to cold tap water source
●
Use the proper inline cartridge filter: Matte finish for the DI intake, chrome finish for the Cold Tap intake