Operating instructions

Baskets and inserts
To ensure that the baskets and inserts
are functioning properly, they must be
checked daily. A checklist is supplied
with the washer-disinfector.
Check the following points:
Are the rollers in proper condition
and are they securely attached to the
basket / insert?
Is the basket connection set to the
correct height and firmly screwed
on?
Are all spray jets, spray sleeves and
hose adapters securely attached to
the basket / insert?
Are all spray jets, spray sleeves and
hose adapters unclogged so that
wash water can flow through?
Are all caps, covers and fasteners
securely attached to the spray
sleeves?
For mobile units in the modular
system, are the caps in the module
connection working properly?
if applicable:
Do the spray arms rotate freely?
Are the spray jets clogged? See the
section on "Cleaning the spray arms".
Machine care
48