User`s manual

6
mm/hr. Formula: mm ÷ h = mm/hr For example: The medication in the syringe is 9
ml. Load this syringe onto the pump, the scale reading of the medication in the syringe
is 40 mm in length. The desired infusion time is 5 hours. 40 mm ÷ 5 hr = 8 mm/hr.
Therefore, set the driving rate to be: 0 8
4) Start Infusion
Turn on the ON/OFF switch to start infusion. The running indicator will flash during
normal operation while the rate setting key is locked. To change the infusion rate, you MUST
reboot the pump. Turn off the ON/OFF switch, change the rate and turn on again.
When batteries begin to run down, an audible and visual alarm is activated. When the
battery voltage cannot sustain a normal operation, LOW BATT and COMPLETE/OCCLUSION
audible and visual alarm will both be activated. When infusion is completed or an occlusion
happens, the pump stops operation with an audible and visual alarm. To continue infusion,
you MUST correct the error and reboot the pump.
5) Patient Controlled ANALGESIA (PCA)
The pump has a PCA feature. As directed by the DVM, if the PCA feature is required,
plug the bolus cable into the pump socket to enable the PCA feature. During a normal infusion,
while additional injections may be needed, you can push the button on the bolus cable to inject
an additional dosage. Each PCA dosage is 1/4 of the preset delivery dosage. For the safety of
the patient and to control the frequency of PCA injections, the interval between two PCA
injections is programmed to be no less than 30 minutes.
STORAGE
This product should be stored in an environment with ambient temperature of 14˚F-131˚F
(-10
- +55 ), relative humidity below 93%, and air pressure 500hpa~1060hpa. Do not store
close to corrosive vapour and/or harmful impurities.