User`s manual
3
TECHNICAL SPECIFICATIONS
OPERATION PROCEDURES
1) Signs and Symbols
a) Type of Protection Against Electric Shock: Typ e : BF
b) Degree of Protection Against Ingress of Fluids: Drip-Proof: IPX1
c) Exclamatory mark
: CAUTION! Please refer to user’s manual.
Applicable syringes
2ml - 50 / 60ml
*
disposable syringes
Linear Flow Rate 1 mm/hr - 99 mm/hr (linear rate)
50/60 ml syringes 0.7 up to 65 ml/hr
20ml syringes 0.35 up to 34 ml/hr
10 ml syringes 0.1 up to 16 ml/hr
5ml syringes 0.1 up to 11 ml/hr
Accuracy less than ± 2 %
Alarms
Audible and Visual Alarms of "COMPLETE /
OCCLUSION" and "LOW BATT"
P.C.A. Function
(BOLUS)
Easily adding 1/4 of preset dosage by bolus cable
without interruption of current infusion. Minimum
30 minutes interval for safety protection
Included Accessories a) 3 x AA batteries
b) 1 x BOLUS cable (PCA)
c) 2 x Syringe protecting cover
d) 1 x Carrying case
Power Supply DC 4.5V (3 x AA batteries)
Battery life 30 days or minimum 30 syringes
Dimensions W 166 × H 60 × D 30 mm (6.5"W × 2.4"H ×1.2"D)
Weight 180 g / 0.55 lb (240g with 3 AA batteries)
* 50/60ml syringe can fill up to 33ml
Flow Rate
(Converted based on
syringe sizes)