Versaflo™ 100 Syringe Pump User’s Manual For Veterinary Use Only Midmark Corporation 60 Vista Drive, Box 286 Versailles, OH 45380-0286 Customer Service and Technical Service TEL: 1-800-MIDMARK (1-800-643-6275) www.midmark.
Versaflo™ 100 Portable Syringe Pump Features Complete portable and light weight Cordless design, with convenient carrying case Bolus function - Adding dosage without interruption of current infusion Scratch-proof syringe cover and carrying/mount case Table of Contents PRECAUTIONS ..................................................................................................... 2 APPLICATION FIELDS..........................................................................................
PRECAUTIONS IMPORTANT: USER SHOULD READ THIS ENTIRE MANUAL BEFORE OPERATING THIS PORTABLE SYRINGE PUMP. SAVE THIS MANUAL FOR YOUR FUTURE REFRENCE! WARNINGS: CRITICAL! DO NOT USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS! DO NOT USE A PUMP, SYRINGE, OR ACCESSORY THAT SHOWS ANY SIGN OF DAMAGE. ONLY AUTHORIZED SERVICE PERSONNEL SHOULD REPAIR THIS PUMP. MANUFACTURER ASSUMES NO RESPONSIBILITY FOR INCIDENTS THAT OCCUR IF THE PUMP IS NOT REPAIRED ACCORDING TO MANUFACTURER-AUTHORIZED PROCEDURES.
TECHNICAL SPECIFICATIONS Applicable syringes 2ml - 50 / 60ml* disposable syringes Linear Flow Rate 1 mm/hr - 99 mm/hr (linear rate) Flow Rate (Converted based on syringe sizes) 50/60 ml syringes 0.7 up to 65 ml/hr 20ml syringes 0.35 up to 34 ml/hr 10 ml syringes 0.1 up to 16 ml/hr 5ml syringes 0.1 up to 11 ml/hr Accuracy less than ± 2 % Alarms Audible and Visual Alarms of "COMPLETE / OCCLUSION" and "LOW BATT" P.C.A.
Keypad Descriptions 2) Load Syringe Fill the syringe with desired medication. Load the syringe onto the pump. (Refer to Figure 2.) Place the syringe into the bed. Be sure to insert the nozzle into the syringe slot. Then press and hold the secure button on the driver to bring the driver block into such a position that the extremity of the syringe plunger can be placed into the driver slot. Fasten the rubber strip to secure the syringe.
3. Move the syringe by holding the syringe driver head by 1 ml according to the reading on the syringe. 4. Note the new reading of the pump’s cursor location. (E.g. 10mm on the pump scale) 5. Calculate the difference on the pump’s scale reading. (E.g. 10mm – 6mm = 4 mm). The difference is the conversion factor between mm and ml of the current loaded syringe per ml. (E.g. 4 mm = 1 ml, the conversion factor is 4 mm/ml) 6.
mm/hr. Formula: mm ÷ h = mm/hr For example: The medication in the syringe is 9 ml. Load this syringe onto the pump, the scale reading of the medication in the syringe is 40 mm in length. The desired infusion time is 5 hours. 40 mm ÷ 5 hr = 8 mm/hr. Therefore, set the driving rate to be: 4) 0 8 Start Infusion Turn on the ON/OFF switch to start infusion. The running indicator will flash during normal operation while the rate setting key is locked. To change the infusion rate, you MUST reboot the pump.
Versaflo 100 syringe pump can be mounted on the cage using the carrying case included in the package LIMITED WARRANTY The Syringe Pump has been carefully manufactured from the highest quality components. The pump is guaranteed against defects in material and workmanship for two (2) years from date of purchase by the original purchaser.
This Limited Warranty shall not apply to defects or damage caused, wholly or in part, by negligence, spilt fluids, dropping of the pump, misuse, abuse, improper installation or alteration by anyone other than qualified, trained personnel; or to damage resulting from inadequate packaging in returning the pump.