Calling for Service
Contact your Midmark Authorised Dealer, or log on to www.midmark.com. Model and serial number
information will be required when calling for service and can be found on the back of the unit.
To contact Midmark directly:
1-800-Midmark (1-800-643-6275) or 937-526-3662
8:00 am to 5:00 pm EST (Monday thru Friday)
(excluding standard U.S. holidays)
ABORT (unit over temp)
If the unit overheats due to prolonged use, the following screens will scroll.
Solutions:
The QuickClean™ has detected that the
ultrasonic generator on the unit has
operated above the rated temperature and
has disabled the ultrasonic function.
This also disables a cycle from starting.
The most likely cause is the liquid
evaporating when the level sensor
is switched off.
Switch the unit off, allow the unit to cool for
a few minutes, top up the liquid above the
level sensor and switch the unit back on.