ABORT OVER TEMP
If the liquid temperature exceeds the allowed maximum of 140°F (60°C), the following
screens will scroll.
Solutions:
• The QuickClean™ has detected that the
water in the bath is above 140ºF/60ºC and
has aborted the cycle.
• Please allow the liquid to cool or drain some
fluid and replace with cooler liquid.
ABORT SONICS LOW
If the power to the ultrasonic generator should fall below a certain level during operation,
the cleaning cycle will stop and the following screens will alternate.
Solutions:
• The QuickClean™ is fitted with an ultrasonic power detector, which monitors the ultrasonic activ-
ity in the bath during a cleaning cycle.
• If the ultrasonic power drops below a certain point, the cycle will be incomplete and this message
will display. This is not necessarily caused by a problem with the machine.
• If this message displays, run another cycle to see if the problem persists.