Step 3.) Cut a piece of foil approximately 6” in length and wide enough to attach to the support
prepared in step 2 and suspend almost to the bottom of the tank. The foil must not touch the
bottom or sides of the tanks. Care should be taken to avoid crinkling of the foil. (See photo 1)
Step 4.) With the foil test sample removed from the tank, prepare a fresh ultrasonic cleaner solution by
filling the tank to the fill line. Add the recommended amount of Midmark cleaning solution and
finish filling the tank to specified level with water.
Step 5.) Turn the ultrasonic on and run the degas cycle 10 minutes to degas the solution.
Step 6.) With the ultrasonic cleaner off and the basket removed, lower the foil sample into the center
of the tank positioning it so the foil isn’t touching the sides or bottom of the tank. For larger
units, e.g. QuickClean™ QC6, multiple samples should be used to get a better sampling of the
uniformity of the ultrasonic cavitation in the tank. (See photo 2)