Guidance and Manufacture’s Declaration – Electromagnetic Immunity
The Ultrasonic Cleaning Bath is intended for use in the electromagnetic environment speci ed below. The
customer or the user of the Ultrasonic Cleaning Bath should ensure that it is used in such an environment
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environmental -
guidance
Electrostatic
Discharge
IEC 61000-4-2:2008
±8.0 kV Contact
±4.0 kV,
±8.0 kV, ±15.0 kV
Air
±8.0 kV Contact
±4.0 kV,
±8.0 kV, ±15.0 kV
Air
The fl oor should be in wood, concrete
or ceramic tile.
If the fl oor is covered with synthetic
material, relative humidity should be at
least 30%.
EFT/Bursts
IEC 61000-4-4:2012
0.5 kV, ±1.0 kV,
±2.0kV Live,
Neutral and Earth
0.5 kV, ±1.0 kV,
±2.0kV Live,
Neutral and Earth
The quality of the power supply should
be that one of a typical commercial or
clinical environment.
Surge Immunity
IEC 61000-4-5:2005
±0.5 kV, ±1.0 kV
Line to Line ±0.5
Kv, ±1.0 kV, ±2.0 kV
Line to Earth
±0.5 kV, ±1.0 kV
Line to Line ±0.5
Kv, ±1.0 kV, ±2.0 kV
Line to Earth
The quality of the power supply should
be that one of a typical commercial or
clinical environment.
Magnetic Field
Immunity
IEC 61000-4-8:2009
30A/m 30A/m Power frequency magnetic elds
should have characteristic levels
of a typical commercial or clinical
environment.
Dips & Interruptions
EN 61000-4-11:2004
0% for 10 ms at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315
0% for 10 ms at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315
The quality of the power supply should
be that one of a typical commercial
or clinical environment. If the user
requests a continuous functioning
during power supply interruptions, it is
recommended a backup power supply
unit (UPS) or battery.