Guidance and Manufacture’s Declaration – Electromagnetic Immunity
The Ultrasonic Cleaning Bath is intended for use in the electromagnetic environment speci ed below. The
customer or the user of the Ultrasonic Cleaning Bath should ensure that it is used in such an environment
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environmental -
guidance
Electrostatic
Discharge
IEC 61000-4-2:2008
±8.0 kV Contact
±4.0 kV,
±8.0 kV, ±15.0 kV
Air
±8.0 kV Contact
±4.0 kV,
±8.0 kV, ±15.0 kV
Air
The fl oor should be in wood, concrete
or ceramic tile.
If the fl oor is covered with synthetic
material, relative humidity should be at
least 30%.
EFT/Bursts
IEC 61000-4-4:2012
0.5 kV, ±1.0 kV,
±2.0kV Live,
Neutral and Earth
0.5 kV, ±1.0 kV,
±2.0kV Live,
Neutral and Earth
The quality of the power supply should
be that one of a typical commercial or
clinical environment.
Surge Immunity
IEC 61000-4-5:2005
±0.5 kV, ±1.0 kV
Line to Line ±0.5
Kv, ±1.0 kV, ±2.0 kV
Line to Earth
±0.5 kV, ±1.0 kV
Line to Line ±0.5
Kv, ±1.0 kV, ±2.0 kV
Line to Earth
The quality of the power supply should
be that one of a typical commercial or
clinical environment.
Magnetic Field
Immunity
IEC 61000-4-8:2009
30A/m 30A/m Power frequency magnetic  elds
should have characteristic levels
of a typical commercial or clinical
environment.
Dips & Interruptions
EN 61000-4-11:2004
0% for 10 ms at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315
0% for 10 ms at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315
The quality of the power supply should
be that one of a typical commercial
or clinical environment. If the user
requests a continuous functioning
during power supply interruptions, it is
recommended a backup power supply
unit (UPS) or battery.