Operating Environment
Ambient Temperature 41°F to 104°F (5 to 40°C)
Maximum relative humidity 80% R.H. in room temperatures up to 88°F (31°C) decreasing linearly
to 50% R.H. at 104°F (40°C)
Altitude above sea level Up to 6,560 ft (2000 m)
Operating Environment Indoor use only
Specifications
UL 61010-1:2012/R:2018-11-21 &
CAN/CSA C22.2 No. 61010-1:2012/
A1:2018-11
Safety requirements for electrical equipment for
measurement, control and laboratory use
FCC CFR 47 Part 15 PART B EMC Compliance
IEC 60601-1-2:2014 EMC Compliance
(EC medical devices emissions Group 1 Class A)
IEC 61326:2012 EMC compliance
(EC, emissions Group 1 Class A)
Electrical Requirements
Standards
NOTE
: If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired
Connect to a 115 VAC 60 Hz fully grounded supply via a 3 pin plug. It can be dangerous to operate an
ultrasonic bath without a grounded connection.
The power plug is fitted with a 10A fuse (T10A H 250V). NEVER FIT A FUSE OF A HIGHER RATING.
Ensure that excess power cable is stored neatly.
115 Electrical Details
Power supply: 115V @ 60 Hz
Ultrasonic frequency: 32 to 38kHz
Pollution degree: 2
Installation category: II
Note: Supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage