Cycle Count
The cycle count shows all cycles which the bath has begun and includes those cycles which are not
completed; i.e. those cycles aborted by the operator or because the liquid level was low.
This example shows the number of cycles which the QuickClean™ ultrasonic bath has run since new.
Low Power
The QuickClean™ can be set to save power when not in use.
In its factory preset mode, ON means that after five minutes of inactivity, the QuickClean™ ultra-
sonic bath will enter sleep mode, the screen illumination will power down.
The QuickClean™ uses less than 10 Watts of power while in sleep mode.
By setting Low Power to OFF, the QuickClean™ screen will be illuminated constantly while the
machine is plugged in.
Liquid Level Sensor
The liquid level sensor is preset at the factory to ON. Meaning if the liquid level falls below the
required level during operation, the cycle will STOP.
Turning the level sensor OFF is for special cleaning applications with very low conductivity liquids
such as pure deionized (DI) water.