User Guide

Table Of Contents
English - 12
003-10007-99 © Midmark Corporation 2017
Preparing to Run a Cycle
After the degas cycle, or if the degas cycle was skipped the screens above
will appear alternately.
A) Place items to be cleaned in the basket.
B) Insert accessory basket into the ultrasonic cleaner.
C) Check the bath for the required level of water and cleaning solution.
D) Place the lid on the unit prior to starting the cycle.
E) The ultrasonic is now ready for use. Press the
left arrow key “START” to begin
the cleaning cycle.
Note:
The ultrasonic has been pre programmed with a cleaning cycle time of 10
minutes. Some components may take longer or shorter to clean effectively
depending on the contaminants and the level of contamination. To adjust
the cycle time refer to Adjusting cycle time section.
Caution
Always rinse and inspect your instruments to ensure all debris has been removed prior to
sterilization.
Equipment Alert
If you are using your bath to clean medical equipment, it is recommended to limit
the liquid temperature to 55°C (130°F). This will avoid the coagulation of protein and
other bio burden which can then bake onto the equipment.