Specifications

SECTION V
SCHEMATICS AND DIAGRAMS
Figure 5-3. (Sheet 1 of 3) M11 Wiring Diagram
(Applies to Units With Serial Numbers ES-1000 Thru ES-1346 And ET-1000 Thru ET1094)
© Midmark Corporation 2002 SF-1827 Rev. 5/03 Page 5-6 Printed in U.S.A.
1
2
3
4
5
Printer
Printer
Connector
12345
Display Panel
PC Board
Ribbon Cable
Connector
Orange (Ground)
Green (Ground)
Red (Power)
Blue (Busy)
White (Data)
PCB/Printer
Harness
Temperature
Probe Assembly
Level
Sensor Assembly
Blue
Door Switch
Yellow
Green
PCB/Heater
Harness
Heating Element
N.O.
Control
PC Panel
PCB/AC/Thermostat
Harness (015-0877-00)
Black
White
Green/Yellow or
Green
Overheat
thermostat
Cordset
A.C. Connector
Receptacle
Green
Fill Solenoid
Vent Solenoid
Black
Black
Red
Red
Red
Black
Chamber Level
Sensor Harness
Black (+5 V.)
Red (V. Out)
White (Ground)
Flex Guard
Tubing
Hot Neutral
Ground
Black
L
G
N
U.S.A. 015-0640-00
Cont. Europe 015-0688-07
Italian 015-0688-08
Swiss 015-0688-09
United Kingdom 015-0688-10
Australian 015-0688-11
Israeli 015-0688-12
Japanese 015-0363-06
120V Unit 10.00 to 11.05 ohms
230V Unit 38.04 to 42.05 ohms
120V Unit 140 ohms
230V Unit 818 ohms
120V Unit 400 ohms
230V Unit 1960 ohms
Black
Black
Door Pulse
Solenoid
120V Unit 11.2 (+/- 1) ohms
240V Unit 45.0 (+/- 4) ohms
(*) NOTE:
Some Solenoid Coils may be marked
FWR (Full Wave Rectified).
Use the
Mohms
scale to measure these coils.
If reading is 0.2 Mohms or above, coil is
GOOD
. If OL is displayed, coil is
BAD
(*)
MA246708
Return To Table Of Contents