Specifications
SECTION V
SCHEMATICS AND DIAGRAMS
© Midmark Corporation 2002 SF-1827 Rev. 5/03 Page 5-3 Printed in U.S.A.
MA376202
Figure 5-1 (Sheet 3 of 3) M9 Wiring Diagram
(Applies To Serial Numbers CZ4591 thru Present, DA1005 thru Present, DB1150 thru Present,
DX1970 thru Present, DY1139 thru Present, FD1000 thru Present, and OM14724 thru Present)
Display Panel
PC Board
Ribbon Cable
Connector
Temperature
Probe Assembly
Level
Sensor Assembly
Blue
Door Switch
N.O.
Control
PC Panel
Fill Solenoid
Vent Solenoid
Black
Black
Red
Red
Black
Chamber Level
Sensor Harness
Black (+5 V.)
Red (V. Out)
White (Ground)
Flex Guard
Tubing
Hot
Neutral
Ground
Black
1
2
3
4
5
Printer
Printer
Connector
12345
Orange (Ground)
Green (Ground)
Red (Power)
Blue (Busy)
White (Data)
PCB/Printer
Harness
Red
PCB/AC/Thermostat
Harness
Black
White
Green/Yellow
Overheat
thermostat
230 V.
Cordset
A.C. Connector
Receptacle
L
G
N
Yellow
Green
PCB/Heater
Harness
Heating Element
Black
White
Green/Yellow
Overheat
thermostat
015-1138-00
Cont. Europe 015-0688-07
Italian 015-0688-08
Swiss 015-0688-09
United Kingdom 015-0688-10
Australian 015-0688-11
Israeli 015-0688-12
Japanese 015-0363-06
120 V. CORDSET
(U.S.A.)
015-1139-00
100V Unit 6.81 to 7.52 ohms
120V Unit 9.77 to 10.80 ohms
240V Unit 40.46 to 44.71 ohms
100V Unit 137 ohms
120V Unit 140 ohms
240V Unit 818 ohms
100V Unit 375 ohms
120V Unit 400 ohms
240V Unit 1960 ohms
Door Pulse
Solenoid
100V Unit 7.90 ohms
120V Unit 14.0 ohms
240V Unit 52.0 ohms
Black
Black
(*) NOTE:
Some Solenoid Coils may be marked
FWR (Full Wave Rectified).
Use the
Mohms
scale to measure these coils.
If reading is 0.2 Mohms or above, coil is
GOOD
. If OL is displayed, coil is
BAD
(*)
Return To Table Of Contents