Specifications

SECTION V
SCHEMATICS AND DIAGRAMS
MA246707
Figure 5-1 (Sheet 2 of 3) M9 Wiring Diagram
(Applies To Serial Numbers CZ1685 thru CZ4590, DB1011 thru DB1149, OM5131 thru OM14723,
DX1171 thru DX1969, DA1002 thru DA1004, and DY1008 thru DY1138)
© Midmark Corporation 2002 SF-1827 Rev. 5/03 Page 5-2 Printed in U.S.A.
1
2
3
4
5
Printer
Printer
Connector
12345
Display Panel
PC Board
Ribbon Cable
Connector
Orange (Ground)
Green (Ground)
Red (Power)
Blue (Busy)
White (Data)
PCB/Printer
Harness
Temperature
Probe Assembly
Level
Sensor Assembly
Blue
Black
Black
Door Switch
Yellow
Green
PCB/Heater
Harness
Heating Element
N.O.
Control
PC Panel
PCB/AC/Thermostat
Harness (015-0877-00)
Black
White
Green/Yellow
Overheat
thermostat
Cordset
A.C. Connector
Receptacle
Fill Solenoid
Vent Solenoid
Black
Black
Red
Red
Red
Black
Chamber Level
Sensor Harness
Black (+5 V.)
Red (V. Out)
White (Ground)
Flex Guard
Tubing
Hot Neutral
Ground
Black
L
G
N
U.S.A. 015-0640-00
Cont. Europe 015-0688-07
Italian 015-0688-08
Swiss 015-0688-09
United Kingdom 015-0688-10
Australian 015-0688-11
Israeli 015-0688-12
Japanese 015-0363-06
100V Unit 6.81 to 7.52 ohms
120V Unit 9.77 to 10.80 ohms
240V Unit 40.46 to 44.71 ohms
100V Unit 137 ohms
120V Unit 140 ohms
240V Unit 818 ohms
100V Unit 375 ohms
120V Unit 400 ohms
240V Unit 1960 ohms
Door Pulse
Solenoid
100V Unit 7.90 ohms
120V Unit 14.0 ohms
240V Unit 52.0 ohms
(*) NOTE:
Some Solenoid Coils may be marked
FWR (Full Wave Rectified).
Use the
Mohms
scale to measure these coils.
If reading is 0.2 Mohms or above, coil is
GOOD
. If OL is displayed, coil is
BAD
(*)
Return To Table Of Contents