Specifications

SECTION V
SCHEMATICS AND DIAGRAMS
SECTION V
SCHEMATICS AND DIAGRAMS
5.1 Wiring Diagrams
Figure 5-1 (sheets 1 thru 3) illustrate the wiring connec-
tions for the M9 sterilizer.
Figure 5-2 (sheets 1 & 2) illustrate the wiring connec-
tions for the M9D sterilizer.
Figure 5-1 (Sheet 1 of 3). M9 Wiring Diagram
(Applies To Serial Numbers CZ1000 thru CZ1684, DB1000 thru DB1010, OM1000 thru OM5130,
DX1000 thru DX1170, DA1000 thru DA1001, and DY1000 thru DY1007)
MA246706
© Midmark Corporation 2002 SF-1827 Rev. 5/03
Page 5-1 Printed in U.S.A.
Figure 5-3 (sheets 1 thru 3) illustrate the wiring connec-
tions for the M11 sterilizer.
Figure 5-4 (sheets 1 & 2) illustrate the wiring connec-
tions for the M11D sterilizer.
1
2
3
4
5
Printer
Printer
Connector
12345
Display Panel
PC Board
Ribbon Cable
Connector
(015-0646-00)
Orange (Ground)
Green (Ground)
Red (Power)
Blue (Busy)
White (Data)
PCB/Printer
Harness
(015-0668-00)
Temperature
Probe Assembly
Level
Sensor Assembly
Reservoir
Sensor Assembly
Blue
Door Switch
Door Pulse
Solenoid
Yellow
Green
PCB/Heater
Harness
(015-0657-00)
Heating Element
N.O.
Control
PC Panel
PCB/AC/Thermostat
Harness (015-0877-00)
Black
White
Green/Yellow or
Green
Overheat
thermostat
Cordset
A.C. Connector
Receptacle
Green
(015-0679-00)
Fill Solenoid
Vent Solenoid
Black
Black
Red
Red
Red
Black
Reservoir
Sensor
Harness
(015-0655-00)
Green
Red
Black
Chamber Level
Sensor Harness
(015-0654-00)
Black (+5 V.)
Red (V. Out)
White (Ground)
Flex Guard
Tubing
(015-0257-02)
Hot Neutral
Ground
Black
L
G
U
U.S.A. 015-0640-00
Cont. Europe 015-0688-07
Italian 015-0688-08
Swiss 015-0688-09
United Kingdom 015-0688-10
Australian 015-0688-11
Israeli 015-0688-12
Japanese 015-0363-06
100V Unit 6.81 to 7.52 ohms
120V Unit 9.77 to 10.80 ohms
240V Unit 40.46 to 44.71 ohms
100V Unit 137 ohms
120V Unit 140 ohms
240V Unit 818 ohms
100V Unit 375 ohms
120V Unit 400 ohms
240V Unit 1960 ohms
100V Unit 7.90 ohms
120V Unit 14.0 ohms
240V Unit 52.0 ohms
Black
Black
(*) NOTE:
Some Solenoid Coils may be marked
FWR (Full Wave Rectified).
Use the
Mohms
scale to measure these coils.
If reading is 0.2 Mohms or above, coil is
GOOD
. If OL is displayed, coil is
BAD
(*)