SECTION IV
MAINTENANCE / SERVICE
NOTE
The following step applies only to units with old style
control PC board with EPROM version M or before.
Units with old style control PC board with EPROM
version N or after
or
new style control PC board do
not have wire (4) and terminal (5).
(3) Connect wire (4) to terminal (5).
(4) Install hose clamp (1) and sight/drain tube (2) on
condensing tank assembly (3). Tighten hose
clamp.
(5) Install bellows assembly (Refer to para 4.7).
(6) Install back panel (Refer to para 4.5).
(7) Install top cover (Refer to para 4.2).
(8) Turn sterilizer onto its side.
(9) Connect fill tube (1, Figure 4-24) to elbow (2).
(10) Turn sterilizer upright.
(11) Refill condensing tank with distilled water.
4.23 Pressure Relief Valve Removal /
Installation
A. Removal
(1) Remove condensing tank assembly (Refer to
para 4.22).
(2) Disconnect tube (1, Figure 4-26) from
elbow (2).
(3) Remove elbow (2) from pressure relief valve
(3).
(4) Remove pressure relief valve (3) from pressure
vessel (4).
B. Installation
(1) Coat threads of pressure relief valve (3) and
elbow (2) with teflon tape.
(2) Install pressure relief valve (3) in pressure
vessel (4).
(3) Install elbow (2) on pressure relief valve (3).
(4) Connect tube (1) to elbow (2).
(5) Install condensing tank assembly (Refer to
para 4.22).
MA2067-01
1
2
3
4
5
6
7
8
9
10
Figure 4-25. Condensing Tank Assembly
Removal / Installation
© Midmark Corporation 2002 SF-1827 Page 4-19 Printed in U.S.A.
MA2068-01
1
2
3
4
Figure 4-26. Pressure Relief Valve
Removal / Installation
Return To Table Of Contents