Specifications

SECTION IV
MAINTENANCE / SERVICE
NOTE
The following step applies only to units with old style
control PC board with EPROM version M or before.
Units with old style control PC board with EPROM
version N or after
or
new style control PC board do
not have wire (4) and terminal (5).
(9) Disconnect wire (4) from terminal (5).
(10) Disconnect vent tube (6) from fitting (7).
NOTE
Positions of locknut (9) and screws (8) are reversed
on newer units.
(11) Remove screw (8), locknut (9), and condensing
tank assembly (3) from base (10).
B. Installation
NOTE
Fill tube must be lined up with fitting at this time
because it will be impossible to line up after screw
and locknut are installed.
(1) Position condensing tank assembly (3, Figure 4-
25) on base (10) and secure using screw (8) and
locknut (9).
(2) Connect vent tube (6) to fitting (7).
MA2066-01
1
2
Figure 4-24. Condensing Tank Assembly Fill Tube
Disconnection / Connection
1
2
3
4
MA2065-01
4.22 Condensing Tank Assembly
Removal / Installation
A. Removal
(1) Drain water from condensing tank.
(2) Turn sterilizer onto its back.
(3) Disconnect fill tube (1, Figure 4-24) from elbow
(2).
(4) Turn sterilizer upright.
(5) Remove top cover (Refer to para 4.2).
(6) Remove back panel (Refer to para 4.5).
(7) Remove bellows assembly (Refer to para 4.7).
(8) Loosen hose clamp (1, Figure 4-25) and remove
sight/drain tube (2) and hose clamp from
condensing tank assembly (3).
Figure 4-23. Display PC Board
Removal / Installation
© Midmark Corporation 2002 SF-1827 Page 4-18 Printed in U.S.A.
Return To Table Of Contents