Specifications
SECTION IV
MAINTENANCE / SERVICE
(2) Using rubber hammer, tap filter (1) lightly into
pressure vessel (2).
(3) Install wire tray rack and tray plate (Refer to
para 4.17).
4.20 Door Gaskets Removal / Installation
A. Removal
(1) Open sterilizer door.
(2) Remove dam gasket (1, Figure 4-22) and door
gasket (2) from door (3).
B. Installation
(1) Install door gasket (2) in groove in door (3).
NOTE
Some models may have two mounting pins on door
instead of flap. Secure dam gasket by pressing onto
mounting pins.
(2) Install dam gasket (1) by inserting bottom edge
of dam gasket in slot of door gasket (2) and
tucking top edge of dam gasket under flap of
door (3).
(3) Close sterilizer door.
4.21 Display PC Board Removal /
Installation
A. Removal
(1) Remove top cover (Refer to para 4.2).
(2) Remove six screws (1, Figure 4-23) and display
PC board (2) from top cover (3).
(3) If damaged, remove gasket (4) from top
cover (3).
B. Installation
(1) If removed, install new gasket (4) by peeling
protective backing off of gasket and pressing
gasket firmly into position on top cover (3).
(2) Install display PC board (2) on top cover (3) and
secure using six screws (1).
(3) Install top cover (Refer to para 4.2).
MA2063-01
1
2
Figure 4-21. Filter Removal / Installation
© Midmark Corporation 2002 SF-1827 Page 4-17 Printed in U.S.A.
MA2064-01
1
2
3
FLAP
SLOT
GROOVE
Figure 4-22. Door Gaskets Removal / Installation
Return To Table Of Contents