Specifications
SECTION IV
MAINTENANCE / SERVICE
B. Installation
(1) Position door switch (5) in door switch
bracket (4).
(2) Using 3/32 in. punch, install two roll pins (3) in
door switch bracket (4) (Refer to Table 1-2 for
special tool).
(3) Connect wire harness (1) to connector (2).
(4) Install right hand side panel (Refer to
para 4.3).
(5) Close sterilizer door.
4.15 Pulse Solenoid Removal /
Installation
(M9 / M11 only)
A. Removal
(1) Remove control PC board (Refer to para 4.13).
(2) Remove two screws (1, Figure 4-17) from pulse
solenoid (2).
(3) Remove nut (3) and shoulder screw (4), and
then remove pulse solenoid (2) and latch lever
(5) as an assembly from pressure vessel
bracket (6).
(4) Using 3/32 in. punch, remove roll pin (7) and
latch lever (5) from plunger of pulse solenoid (2)
(Refer to Table 1-2 for special tool).
B. Installation
(1) Using 3/32 in. punch, install latch lever (5) on
plunger of pulse solenoid (2) and secure using
roll pin (7) (Refer to Table 1-2 for special tool).
(2) Coat shoulder screw (4) with Loctite 271.
(3) Install latch lever (5) and pulse solenoid (2) as
an assembly on pressure vessel bracket (6).
Secure using shoulder screw (4) and nut (3).
4.14 Door Switch Removal / Installation
A. Removal
NOTE
Units prior to serial number CZ1110/OM1000 have a
different switch weldment. Since this procedure
tampers with the pressure vessel integrity, it should
not be attempted until a Midmark service representa-
tive has been contacted.
(1) Open sterilizer door.
(2) Remove right hand side panel (Refer to para
4.3).
(3) Disconnect wire harness (1, Figure 4-16) from
connector (2).
(4) Using 3/32 in. punch, remove two roll pins (3)
from door switch bracket (4) (Refer to Table 1-2
for special tool).
(5) Remove door switch (5) from door switch
bracket (4).
© Midmark Corporation 2002 SF-1827 Page 4-13 Printed in U.S.A.
Figure 4-16. Door Switch Removal / Installation
2
MA2059-02
4
3
1
5
Return To Table Of Contents