Specifications

SECTION IV
MAINTENANCE / SERVICE
4.12 Fill Solenoid Removal / Installation
NOTE
Steps will vary slightly for M9D / M11D units. Use
this procedure as a guide.
A. Removal
NOTE
Units prior to serial number CZ1110/OM1000 have a
manifold assembly which secures the fill solenoid.
Refer to the parts list to remove the fill solenoid on
these sterilizers.
(1) Drain water from condensing tank.
(2) Remove back panel (Refer to para 4.5).
(3) Tag and disconnect two wires (1, Figure 4-12)
from terminals (3).
(4) Tag and disconnect two wires (2) from
terminals (4).
(5) Disconnect three tubes (5 thru 7) from elbows (8
thru 10).
(6) Remove two screws (11) from bottom of
manifold assembly (12).
NOTE
Spacer (12A) is only on newer units.
(7) Remove manifold assembly (12) and spacer
(12A) from base (13).
(8) Hold screw (1, Figure 4-13) and loosen nut (2).
Turn terminals out of the way.
(9) Remove assembled vent solenoid (3) and elbow
(4) from tee (5).
(10) Remove elbow (6) and tee (5) from fill
solenoid (7).
(11) Remove elbow (8) from fill solenoid (7).
B. Installation
(1) Coat threads of tee (5, Figure 4-13) and elbows
(6 and 8) with teflon tape.
(2) Install elbow (8) on fill solenoid (7).
(3) Install elbow (6) and tee (5) on fill solenoid (7).
(4) Loosen nut (9). Turn terminals of fill solenoid
(7) so wires will connect easily to fill solenoid.
Tighten nut (9).
(5) Coat threads of elbow (4) with teflon tape.
MA2056-0
2
1
2
3
4
5
6
7
8
9
10
11
12
13
12A
Figure 4-12. Fill Solenoid Removal / Installation
© Midmark Corporation 2002 SF-1827 Page 4-9 Printed in U.S.A.
Figure 4-13. Fill Solenoid Fittings
Removal / Installation
1
2
3
4
5
TERMINALS
6
7
8
9
MA205801
Return To Table Of Contents