Specifications
SECTION IV
MAINTENANCE / SERVICE
EQUIPMENT ALERT
When installing the vent solenoid, the
side of the solenoid marked IN needs to
be connected to the pressure line. Reversing vent
solenoid will cause system failure.
(7) Install spacer (12A, Figure 4-10) and manifold
assembly (12) on base (13) and secure using
two screws (11).
(8) Connect three tubes (5 thru 7) to elbows (8 thru
10).
(9) Connect two wires (2) to terminals (4).
(10) Connect two wires (1) to terminals (3).
(11) Install back panel (Refer to para 4.5).
(12) Refill condensing tank with distilled water.
MA2056-0
2
1
2
3
4
5
6
7
8
9
10
11
12
13
12A
Figure 4-10. Vent Solenoid Removal / Installation
B. Installation
NOTE
The end of elbow (6) which receives a tube does not
get coated with teflon tape.
(1) Coat the threads of elbows (4 and 6, Figure 4-
11) with teflon tape.
(2) Install elbows (4 and 6) on vent solenoid (3).
(3) Hold screw (1) and loosen nut (2). Turn termi-
nals out of the way.
(4) Install assembled elbow (4) and vent solenoid
(3) on tee (5).
(5) Turn terminals of vent solenoid (3) so wires will
connect easily to vent solenoid.
(6) Hold screw (1) and tighten nut (2).
© Midmark Corporation 2002 SF-1827 Page 4-8 Printed in U.S.A.
1
2
3
4
5
6
TERMINALS
MA205701
Figure 4-11. Vent Solenoid Fittings
Removal / Installation